Descripción Current Transducer HTA 100~1000-S
Fabricantes LEM 
Logotipo LEM Logotipo



Hay una vista previa y un enlace de descarga de HTA200-S (archivo pdf) en la parte inferior de esta página.


Total 2 Páginas

No Preview Available ! HTA200-S Hoja de datos, Descripción, Manual

Current Transducer HTA 100..1000-S
For the electronic measurement of DC, AC and pulsed currents,
with a galvanic isolation between the primary (high power) circuit
and the secondary (electronic) circuit.
IPN = 100 A
Electrical data
Primary nominal
r.m.s. current
IPN (A)
100
200
300
400
500
600
1000
Primary current
measuring range
IP (A)
± 300
± 600
± 900
± 1000
± 1000
± 1000
± 1000
ÎP Overload capacity (Ampere Turns)
VOUT Analogue output voltage @ ± IPN
RL Load resistance TA = 0 .. + 70°C
TA = - 25 .. + 85°C
VC Supply voltage (± 5 %)
IC Current consumption (max)
Vb Rms rated voltage1)
Vd Rms voltage for AC isolation test, 50 Hz, 1 mn
Ris Isolation resistance @ 500 VDC
Accuracy - Dynamic performance data
εX Accuracy 2) @ IPN, TA = 25°C, @ ± 15 V
L Linearity 2)
VOE
VOM
εVOT
TC G
tr
di/dt
f
Electrical offset voltage @ IP = 0, TA = 25°C
Residual offset voltage @ IP = 0
after an overload of 3 x IPN
Thermal drift of offset voltage TA = - 25 .. + 85°C
Thermal drift of gain TA = - 25 .. + 85°C
Response time @ 90 % of IP
di/dt accurately followed
Frequency bandwidth (- 3 dB) 3)
Type
HTA 100-S
HTA 200-S
HTA 300-S
HTA 400-S
HTA 500-S
HTA 600-S
HTA 1000-S
30000
A
±4 V
> 1 k
> 3 k
± 15 V
25 mA
500 V
3 kV
> 500
M
±1
± 0.5
Max
± 10
± 10
±1
± 0.05
<3
> 50
DC .. 50
%
%
mV
mV
mV/°K
%/°K
µs
A/µs
kHz
General data
TA Ambient operating temperature
TS Ambient storage temperature
m Mass
- 25 .. + 85
- 25 .. + 85
230
°C
°C
g
Standards Safety
EN50178 (1994)
EMC
EN50082-2 (1992)
EN50081-1 (1992)
Deviation in output when tested to EN 61000-4-6
Deviation in output when tested to EN 61000-4-4
< 10
< 10
% of IPN
% of IPN
Notes : 1) Overvoltage Category III, Pollution Degree 2
2) Excludes the electrical offset
3) Refer to derating curves in the technical file to avoid excessive core
heating at high frequency
Features
Open loop transducer using Hall
Effect
Panel mounting - Horizontal or
Vertical
Insulated plastic case to UL 94-V0.
Advantages
Very good linearity
Very good accuracy
Low temperature drift
Wide frequency bandwidth
Very low insertion losses
High immunity to external
interference
Current overload capability
Low power consumption
Wide dynamic range, 100 to 1000 A
in one package.
Applications
AC variable speed drives and servo
motor drives
Static converters for DC motor drives
Battery supplied applications
Uninterruptable Power Supplies
(UPS)
Switched Mode Power Supplies
(SMPS)
Power supplies for welding
applications.
001107/1
LEM Components
www.lem.com

1 page





PáginasTotal 2 Páginas
PDF Descargar[ Datasheet HTA200-S.PDF ]




Hoja de datos destacado

Número de piezaDescripciónFabricantes
HTA200-SCurrent Transducer HTA 100~1000-SLEM
LEM

Número de piezaDescripciónFabricantes
SLA6805M

High Voltage 3 phase Motor Driver IC.

Sanken
Sanken
SDC1742

12- and 14-Bit Hybrid Synchro / Resolver-to-Digital Converters.

Analog Devices
Analog Devices


DataSheet.es es una pagina web que funciona como un repositorio de manuales o hoja de datos de muchos de los productos más populares,
permitiéndote verlos en linea o descargarlos en PDF.


DataSheet.es    |   2020   |  Privacy Policy  |  Contacto  |  Buscar